Validation of a Monte Carlo dose calculation algorithm for clinical electron beams in the presence of phantoms with complex heterogeneities

Download pdf | Full Screen View

Date Issued:
2014
Summary:
The purpose of this thesis is to validate the Monte Carlo algorithm for electron radiotherapy in the Eclipse™ treatment planning system (TPS), and to compare the accuracy of the Electron Monte Carlo algorithm (eMC) to the Pencil Beam algorithm (PB) in Eclipse™. Dose distributions from GafChromic™ EBT3 film measurements were compared to dose distributions from eMC and PB treatment plans. Measurements were obtained with 6MeV, 9MeV, and 12MeV electron beams at various depths. A 1 cm thick solid water template with holes for bone-like and lung-like plugs was used to create assorted configurations and heterogeneities. Dose distributions from eMC plans agreed better with the film measurements based on gamma analysis. Gamma values for eMC were between 83%-99%, whereas gamma values for PB treatment plans were as low as 38.66%. Our results show that using the eMC algorithm will improve dose accuracy in regions with heterogeneities and should be considered over PB.
Title: Validation of a Monte Carlo dose calculation algorithm for clinical electron beams in the presence of phantoms with complex heterogeneities.
228 views
24 downloads
Name(s): Enright, Shayla Landfair, author
Pella, Silvia, Thesis advisor
Florida Atlantic University, Degree grantor
Charles E. Schmidt College of Science
Department of Physics
Type of Resource: text
Genre: Electronic Thesis Or Dissertation
Date Created: 2014
Date Issued: 2014
Publisher: Florida Atlantic University
Place of Publication: Boca Raton, Fla.
Physical Form: application/pdf
Extent: 93 p.
Language(s): English
Summary: The purpose of this thesis is to validate the Monte Carlo algorithm for electron radiotherapy in the Eclipse™ treatment planning system (TPS), and to compare the accuracy of the Electron Monte Carlo algorithm (eMC) to the Pencil Beam algorithm (PB) in Eclipse™. Dose distributions from GafChromic™ EBT3 film measurements were compared to dose distributions from eMC and PB treatment plans. Measurements were obtained with 6MeV, 9MeV, and 12MeV electron beams at various depths. A 1 cm thick solid water template with holes for bone-like and lung-like plugs was used to create assorted configurations and heterogeneities. Dose distributions from eMC plans agreed better with the film measurements based on gamma analysis. Gamma values for eMC were between 83%-99%, whereas gamma values for PB treatment plans were as low as 38.66%. Our results show that using the eMC algorithm will improve dose accuracy in regions with heterogeneities and should be considered over PB.
Identifier: FA00004192 (IID)
Degree granted: Thesis (M.S.)--Florida Atlantic University, 2014.
Collection: FAU Electronic Theses and Dissertations Collection
Note(s): Includes bibliography.
Subject(s): Electron beams -- Therapeutic use
Image guided radiotherapy
Monte Carlo method
Proton beams -- Therapeutic use
Radiation dosimetry
Radiotherapy, High energy
Held by: Florida Atlantic University Libraries
Sublocation: Digital Library
Persistent Link to This Record: http://purl.flvc.org/fau/fd/FA00004192
Use and Reproduction: Copyright © is held by the author, with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Use and Reproduction: http://rightsstatements.org/vocab/InC/1.0/
Host Institution: FAU
Is Part of Series: Florida Atlantic University Digital Library Collections.