Abstract/Description:
Stereotactic Body Radiation Therapy (SBRT) is a modern precision radiation therapy to deliver the dose in 1 to 5 fractions with high target dose conformity, and steep dose gradient towards healthy tissues. The dose delivered is influenced by the leaf width of the MLC, especially in case of SBRT. Treatment plans with high definition (HD) MLC having leaf-width 2.5 mm and normal MLC having leaf-width 5 mm, were compared to quantify dosimetric and radiobiological parameters. Dosimetric parameters conformity indices (CI), gradient indices (GI) and heterogeneity indices (HI) were compared. The radiobiological parameters were evaluated by normal tissue complication probability (NTCP) and tumor control probability (TCP) based on the equivalent uniform dose (EUD). The results show that there is dosimetric and radiobiological merit of the HD MLC over the normal MLC. However, the improvement is not consistent with all the plans and thus further research is required prior to conclusion.
Title: Dosimetric and Radiobiological Comparison of the Effects of High Definition versus Normal Collimation on Treatment Plans for Stereotactic Lung Cancer Radiation Therapy.
35 views
6 downloads
Name(s): Pudasaini, Mukunda Prasad, author
Pella, Silvia, Thesis advisor
Leventouri, Theodora, Thesis advisor
Florida Atlantic University, Degree grantor
Charles E. Schmidt College of Science
Department of Physics
Type of Resource: text
Genre: Electronic Thesis Or Dissertation
Date Created: 2018
Date Issued: 2018
Publisher: Florida Atlantic University
Place of Publication: Boca Raton, Fla.
Physical Form: application/pdf
Extent: 61 p.
Language(s): English
Abstract/Description: Stereotactic Body Radiation Therapy (SBRT) is a modern precision radiation therapy to deliver the dose in 1 to 5 fractions with high target dose conformity, and steep dose gradient towards healthy tissues. The dose delivered is influenced by the leaf width of the MLC, especially in case of SBRT. Treatment plans with high definition (HD) MLC having leaf-width 2.5 mm and normal MLC having leaf-width 5 mm, were compared to quantify dosimetric and radiobiological parameters. Dosimetric parameters conformity indices (CI), gradient indices (GI) and heterogeneity indices (HI) were compared. The radiobiological parameters were evaluated by normal tissue complication probability (NTCP) and tumor control probability (TCP) based on the equivalent uniform dose (EUD). The results show that there is dosimetric and radiobiological merit of the HD MLC over the normal MLC. However, the improvement is not consistent with all the plans and thus further research is required prior to conclusion.
Identifier: FA00013148 (IID)
Degree granted: Thesis (M.S.)--Florida Atlantic University, 2018.
Collection: FAU Electronic Theses and Dissertations Collection
Note(s): Includes bibliography.
Subject(s): Radiosurgery
Lung Neoplasms
Radiation dosimetry
Radiobiology
Held by: Florida Atlantic University Libraries
Sublocation: Digital Library
Persistent Link to This Record: http://purl.flvc.org/fau/fd/FA00013148
Use and Reproduction: Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder.
Use and Reproduction: http://rightsstatements.org/vocab/InC/1.0/
Host Institution: FAU
Is Part of Series: Florida Atlantic University Digital Library Collections.